ISO 10360-5-2020 pdf free.Geometrical product specifications(GPS)—Acceptance and reverificationtests for coordinate measuring systems (CMS).
The following documents are referred to in the text in such a way that some or all of their contentconstitutes requirements of this document. For dated references, only the edition cited applies. Forundated references, the latest edition of the referenced document(including any amendments) applies.
rated operating condition:operating condition that needs to be fulfilled during measurement in order for a measuring instrumentor measuring system to perform as designed
Note 1 to entry: Rated operating conditions generally specify intervals of values for a quantity being measuredand for any influence quantity.
Note 2 to entry: Within the IS0 10360 series, the term “as designed” means as specified by MPEs.
Note 3 to entry: lf an MPE specification is thought of as a function (where different MPE values could be given fordifferent conditions), then the rated operating conditions define the domain of that function.
[SOURCE:ISO/IEC Guide 99:2007,4.9, modified —Notes 2 and 3 to entry added.]3.2
inferred qualification:probing system qualification method where the parameters for each probing system attached toan articulation system are inferred by interpolation, extrapolation or another relevant model, forsignificantly different angular position(s) from parameters acquired by empirical qualification (3.3) at afew angular positions.
empirical qualification:probing system qualification method where the parameters for each probing system attached to anarticulation system need to be acquired by measurement of the reference sphere at each angularposition used.
effective diameter:stylus diameter used with the tip correction vector, for compensating stylus centre points to obtainsurface points.
Tacterror of indication of the diameter of a test sphere, the measurements being performed by a CMM witha single stylus, using the discrete-point probing mode.scanning mode form error on a spherePForm.Sph.Scan:k:Tact observed form of a test sphere, the measurements being performed by a CMM with a single stylus,usingscanning mode.
Note 1 to entry: Wherekis replaced by the following designates as applicable: k= PPor NPP depending on systemscanning mode, pre-defined path (PP) or not pre-defined path(NPP).scanning mode size error on a spherePsize.Sph.Scan:k:Tact error of indication of the size of a test sphere, the measurements being performed by a CMM with asingle stylus,using scanning mode.ISO 10360 pdf download.