AS 3825:2020 pdf free.Procedures and devices for the removal, containment and disposal of scalpel blades from scalpel handles.
4 Design requirements for scalpel blade removal devices
The design of the scalpel blade removal device shall
a allow for the safe removal of a scalpel blade with a single-handed operation, (test in accordance with Appendix A);
NOTE 1 This may require the fixing of the scalpel blade removal device to a surface with a bracket or suitable alternative fixation device, e.g. suction cup, reusable adhesive or hook-and-loop fastener.
(b) indicate that the used scalpel blade has been removed by
(i) an audible sound; and/or
(ii) a visual cue;
(c) allow for the safe containment of the scalpel blade immediately after the scalpel blade is removed, (test in accordance with Appendix A);
(d) ensure that the scalpel blade remains completely contained within the device after removal (test in accordance with Appendix B);
(e) be such that during use the loaded scalpel handle is not directed towards the user’s body; and
(f) if a single-use device, allow for accurate count of the contained scalpel blades.
The design of the scalpel blade removal device should also —
(1) permit the removal of the scalpel blade —
(A) with minimum droplet dispersal;
(B) without breakage or flip-back of the blade, and
(C) without damage to the scalpel blade handle; (ii) not require excessive force to remove the scalpel blade;
(iii) if a single-use device, be capable of being sterilized, where claimed by the manufacturer; and
(iv) if a multiple-use device, allow for ease of cleaning of the external surfaces.
NOTE 2 Appendix C may be useful for reporting purposes.
5 Procedures for the removal of scalpel blades
5.1 Location of scalpel blade removal device
The scalpel blade removal device shall be situated as close as possible to the point of use.
5.2 Type of scalpel blade removal device
Where a scalpel blade removal device is loaded onto a reusable handle, removal of the scalpel blade from the handle shall be done with a passive scalpel blade removal device in preference to an active scalpel blade removal device.
5.3 Removal by hand
Scalpel blades shall not be removed by hand (see Figure 1).AS 3825 pdf download.

Download infomation Go to download
Note: Can you help me share this website on your Facebook or others? Many thanks!
AS/NZS 60079.5:2015 pdf free Free AS Standards

AS/NZS 60079.5:2015 pdf free

AS/NZS 60079.5:2015 pdf free.Explosive atmospheres Part 5: Equipment protection by powdered filling 'q'. 4.1.1.3 Containers intended to be opened for repair Electrical equipment, parts of electrical equipment, or Ex components that are designed to be repaired shall...
Get More
AS/NZS 60079.2:2015 pdf free Free AS Standards

AS/NZS 60079.2:2015 pdf free

AS/NZS 60079.2:2015 pdf free.AUSTRALIAN/NEW ZEAL AND STANDARD Explosive atmospheres. 5 Constructional requirements for pressurized enclosures 5.1 Enclosure The pressurized enclosure shall have a degree of protection in accordance with Table 2. For Level of Protection pxb with...
Get More
AS/NZS 5263.1.3:2016 pdf free Free AS Standards

AS/NZS 5263.1.3:2016 pdf free

AS/NZS 5263.1.3:2016 pdf free.Gas appliances Part 1.3: Gas space heating appliances. 2.1 SCOPE This Clause of AS 5263.0 applies. 2.2 GENERAL DESIGN REQUIREMENTS This Clause of AS 5263.0 applies, except as modified below. 2.2.13 IJesign olappliance base...
Get More

LEAVE A REPLY

Anonymous netizen Fill in information