AGMA 935-A05 pdf free.Recommendations Relative to the Evaluation of Radial Composite Gear Double Flank Testers.
This Infomiation sheet provides qualification procedures for double flank testers that are used for the evaluation of radial composite deviations of gears. Recommendations are inckided for establishment of a proper environment Suggested artifact criteria arid analysis procedures are also provided.
2 References
The following documents contain provisions which, through reference in this text may constitute recommendations of this information sheet At the time of publication, the editions re valid. AJI publications are subject to revision, and the users of this information sheet are encouraged to investigate the possality of applying the most recent editions of the publications listed.
AGMA 915-2-A05, inspection Practices – Part 2:
Cylindrical Gears – Radial Measurements
ANSI B89.6.2:2003, Temperature and Humidity Environment for Dimensional Measurement
ANSIIAGMA 101 2-G05, Gear Nomenclature, Definitions of Terms with Symbols
ANSIIAGMA 2015.-i-AOl, Accuracy Classification
System – Tangential Measurements for Cylindrical Gears
3 Definition of terms
The symbols used in this document are shown in table 1.
4 Instrument environment
4.1 Environment
The stability of the environment will affect accuracy of the calibration process and measurement of production parts. The measurement temperature should be maintained as a constant It is recommended that the temperature be 20C. Standards often require an environment controlled to the extent necessary to assure continued measurements of required accuracy considering temperature, humidity, vibration, cleanliness and other controlable factors affecting precision measurement GeneralI environmental effects should not exceed 10% of the desired measuring machine accuracy.
4.1.1 Important parameters
Parameters of primary Importance2):
– the cooling (heating) medium, usually air;
– How rate, distribution and velocity of the cooling (heating) medium;
– frequency and amplitude of temperature variations of the cooling (heating) medium;
– temperature gradients withm the coding (heating) mecHum;
– vibrations;
– electrical power supply quality.
4.1.2 Practical guidelines
The following are practical guidelines for gear measurement However, compliance with these guidelines does not guarantee measurements to a specific accuracy.
– Artifact temperature. Tooling and artifacts should be leftforan adequate perlodto stabilize to ambient temperature. Artifact temperature ideally should be the temperature at which it was calibrated.AGMA 935 pdf free download.
AGMA 935-A05 pdf free