5 Apparatus
5.1 Rigid, flat, horizontal test plane, with coefficient of friction of greater than 0,6 measured by the
method specified in ISO 7176-13 and of sufficient size to conduct the tests. The test plane shall have
a surface that lies between two imaginary horizontal planes 20 mm apart and has no more than 0,5 0
variation in slope or cross slope throughout the test.
The test plane shall be long enough to allow the wheelchair to reach maximum speed.
NOTE An area of approximately 10 m x 3 m is normally of sufficient size but the testing of larger and/or faster wheelchairs might need a larger test plane.
5.2 Rigid, flat, inclined area, with run up and run down, shall be long enough to allow the wheelchair to reach maximum speed and come to a stop within the stated inclination.
The test area of the ramp shall be long enough to allow the wheelchair to come to a stop within the stated inclination which shall be within the tolerance of ±10.
The testing area of the ramp shall have a surface that lies between two imaginary parallel planes 50 mm apart.
The test ramp shall have a coefficient 0 friction in the test area that meets the requirements o
ISO 7 176-13.
The test area shall run immediately from the horizontal test plane via a transition with a radius of less than 25 mm.
The traversed surface shall be free of loose material and steps greater than 5 mm in height.
A ramp approximately 10 m * 3 m is normally of sufficient area, but the testing of larger and/or faster wheelchairs may need a larger ramp.
NOTE 1 The ramp can be of a variable angle type or individual ramps for each required angle.
NOTE 2 Run up and run down areas on either end of the test area need not meet the above criteria. NOTE 3 Figure 1 provides a recommended configuration for the test area and transition.
NOTE 4 A ramp of approximately 10 m x 3 m is normally of sufficient area but the testing of larger and/or faster wheelchairs might need a larger ramp.AS/NZS ISO 7176.2 pdf download.

Download infomation Go to download
Note: Can you help me share this website on your Facebook or others? Many thanks!
AS/NZS 60079.5:2015 pdf free Free AS Standards

AS/NZS 60079.5:2015 pdf free

AS/NZS 60079.5:2015 pdf free.Explosive atmospheres Part 5: Equipment protection by powdered filling 'q'. 4.1.1.3 Containers intended to be opened for repair Electrical equipment, parts of electrical equipment, or Ex components that are designed to be repaired shall...
Get More
AS/NZS 60079.2:2015 pdf free Free AS Standards

AS/NZS 60079.2:2015 pdf free

AS/NZS 60079.2:2015 pdf free.AUSTRALIAN/NEW ZEAL AND STANDARD Explosive atmospheres. 5 Constructional requirements for pressurized enclosures 5.1 Enclosure The pressurized enclosure shall have a degree of protection in accordance with Table 2. For Level of Protection pxb with...
Get More
AS/NZS 5263.1.3:2016 pdf free Free AS Standards

AS/NZS 5263.1.3:2016 pdf free

AS/NZS 5263.1.3:2016 pdf free.Gas appliances Part 1.3: Gas space heating appliances. 2.1 SCOPE This Clause of AS 5263.0 applies. 2.2 GENERAL DESIGN REQUIREMENTS This Clause of AS 5263.0 applies, except as modified below. 2.2.13 IJesign olappliance base...
Get More

LEAVE A REPLY

Anonymous netizen Fill in information