AS 5330:2019 pdf free download.Drying cabinets for reusable medical devices.
Adjustable support brackets, adjustable feet, and levelling bolts fitted to the framework shall be of stainless steel or other suitable material that is corrosion resistant.
1.6 External panels
Panels that are either external, built-in or otherwise not visible, shall be constructed of stainless steel or other suitable material that is corrosion resistant and impervious to moisture.
1.7 Doors and chambers
Doors shall be provided with a catch to keep the doors closed and the chamber airtight. If the system is a pass through, a mechanism shall be put in place to prevent both doors being opened at the same time.
Doors shall be sliding, or hinged to open outwards.
Doors shall be capable of being placed in a fully open or fully closed position, and remain in that position once placed.
Doors may be fitted with glass viewing panels.
Welded sheets used in the construction of the drying cabinet shall not be buckled. Internal welded joints of the chamber shall be smooth, without sharp corners.
The chamber walls and floor shall be concave to prevent liquid draining or spilling through the cabinet door.
1.8 Gaskets
There shall be one gasket [or each door. The gasket shall be designed to produce a cushion effect and to ensure an airtight seal when the door is closed. The gasket shall be mechanically secured and replaceable.
Under operating conditions, gaskets shall not release toxic products and shall fully seal the joint. Door gaskets shall be manufactured from a material suitable for the temperature(s) of operation.
1.9 Air filter
A filter shall be provided to filter incoming air before it enters the chamber for the removal of coarse contaminants such as dust, fluff, lint, and other particulates. It shall be a dry, washable media type filter conforming with AS 1324.1.
The filter shall be accessible and removable for cleaning or replacement.
1.10 Vent
A vent shall be provided for the escape of hot, moist air. It shall be at least 100 mm diameter or an equivalent cross-sectional area. The vent shall be suitable for connection to an exhaust system.
A circulating fan shall be provided, which shall be balanced and removable to allow for cleaning and maintenance of the impeller and the interior of the fan casing.
1.11 Valves
The material for valves shall be suitable for the temperature(s) and pressure(s) of operation. Where non-ferrous metallic valves are fitted, they shall be in accordance with AS 1271.AS 5330 pdf download.

Download infomation Go to download
Note: Can you help me share this website on your Facebook or others? Many thanks!
AS/NZS 60079.5:2015 pdf free Free AS Standards

AS/NZS 60079.5:2015 pdf free

AS/NZS 60079.5:2015 pdf free.Explosive atmospheres Part 5: Equipment protection by powdered filling 'q'. 4.1.1.3 Containers intended to be opened for repair Electrical equipment, parts of electrical equipment, or Ex components that are designed to be repaired shall...
Get More
AS/NZS 60079.2:2015 pdf free Free AS Standards

AS/NZS 60079.2:2015 pdf free

AS/NZS 60079.2:2015 pdf free.AUSTRALIAN/NEW ZEAL AND STANDARD Explosive atmospheres. 5 Constructional requirements for pressurized enclosures 5.1 Enclosure The pressurized enclosure shall have a degree of protection in accordance with Table 2. For Level of Protection pxb with...
Get More
AS/NZS 5263.1.3:2016 pdf free Free AS Standards

AS/NZS 5263.1.3:2016 pdf free

AS/NZS 5263.1.3:2016 pdf free.Gas appliances Part 1.3: Gas space heating appliances. 2.1 SCOPE This Clause of AS 5263.0 applies. 2.2 GENERAL DESIGN REQUIREMENTS This Clause of AS 5263.0 applies, except as modified below. 2.2.13 IJesign olappliance base...
Get More

LEAVE A REPLY

Anonymous netizen Fill in information